Test Code PBDV Lead, Blood, Venous
Specimen Collection Requirements
| Name: | Blood Lead, Venous |
| Alt name: | Blood Lead, Lead |
| Dept: | Elmgrove Chemistry |
| Specimen: | Whole Blood |
| Collection Cont.: | Trace Element Royal blue top tubes (preferred) |
| Also acceptable: | Tan top tubes, Lavender EDTA tube |
| Transport temp: | Refrigerated Temperature |
| Stability: | Refrigerated up to 7 days |
| Specimen Volume: | 1.0 mL (minimum 0.5 mL) |
TECHNICAL SPECIFICATIONS
| Performed: | Business Days |
| Reported: | Next Business Day |
| Methodology: | Inductively Coupled Plasma Mass Spectrometry (ICPMS) |
| Ref range: | Normal: ≤4.9 ug/dL |
| Critcal value: | Critical: ≥45.0 ug/dL |
| Linearity range: | <1.0 to >60.0 ug/dL |
| CPT: | 83655 |
| LOINC: | 77307-7 |
Notes
| The use of a Lavender top tube might produce a falsely elevated blood lead result. |
| It is recommended to redraw in a Trace Element Royal blue top tube. |
| To minimize Lead Contamination: |
| 1. Keep the collection area clean and dust free. |
| 2. Clean the venipuncture site with alcohol, not iodine. |
| 3. Do not ream out or remove blood clots from collection tube. |
| 4. Do not insert any objects into the specimen tube or pour off the sample. |